Displaying reports 1781-1800 of 88643.Go to page Start 86 87 88 89 90 91 92 93 94 End
Reports until 17:47, Friday 06 March 2026
H1 TCS
sophie.muusse@LIGO.ORG - posted 17:47, Friday 06 March 2026 (89416)
CHETA ELECTRONICS

S. Muusse, T. Roocke


All chassis were put on the rack for the Y table and cables attached. All electronics were tested and working however, we have realised additional power supplies are required and some additional cables are required.

Power supply/ cables still required:

A potential solution is to use pre designed 15V, 12V, 5V, regulator boards in another chassis which can be input to the table through the blank and power feedthroughs.  Alternatively there are some preexisting supply boxes for 12 and 15 V

H1 General
ryan.crouch@LIGO.ORG - posted 16:18, Friday 06 March 2026 (89393)
OPS Friday Day shift summary

TITLE: 03/07 Day Shift: 1530-0030 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: Ryan S
SHIFT SUMMARY: We ran into issues trying to lock the IMC after locking JAC and GREEN_ARMs, it seems to be from MC1 (T2 and T3 mostly). 
LOG:                                                                                                                                                                                                                                                                                                                                                                                        

Start Time System Name Location Lazer_Haz Task Time End
15:37 FAC Contractor MidY N Unloading bricks for CEBEX 17:21
15:38 FAC Randy EX N Bring ergo lift back to CS High bay, 15:57 17:34
16:12 FAC Kim, Nellie LVEA N Tech clean, Nellie out 16:48 16:58
16:28 SEI Jeff Optics lab LOCAL SPI work 22:45
17:01 PSL Sheila LVEA N Unlock rotation stage to turn up to 1W then relock, in @ 17:09 17:18
17:19 ALS Sheila LVEA HAM1 Y/LOCAL SGH table alignment work 17:51
17:34 FAC Randy High bay N Clean lift 17:48
17:44 TCS Sophie Cheeta lab N Plug stuff in 17:57
18:29 FAC Contractor MidY N Unload bricks 21:29
18:54 VAC Jordan LVEA N Check HAM1 aux cart 19:07
18:56 TCS Sophie Cheeta / Prep lab N Cheeta work 20:22
19:07 EE Marc LVEA, HAM1 N Terminate cables 22:20
20:16 CDS Dave MER N DAQ chassis work 21:31
21:25 TCS Sophie Prep lab N Cheeta work 22:04
21:48 ISC Masayuki LVEA N Turn on whitening 22:09
22:05 TCS Sophie, Thomas Prep lab N Cheeta 23:30
22:34 EE Marc CER N See about swaping HAM2 MC1 M1 coil driver 22:38
22:47 EE Marc CER N Coil driver swap MC1, in at 22:55 23:25
22:56 EPO Erik + family LVEA N Quick tour 23:13

16:37 EX sensor correction off as the BRS was ringing up from Randys' loading of the ergo lift.

17:18 Sheila unlocked the rotation stage and I turned up the power to 1W, Sheila then relocked it.

17:35 UTC I used the ALS nodes to run an alignment for GRREN_ARMs, ALSX locked after a few taps of ETMX but ALSY took more convincing

19:30 UTC I started to run TFs for MC1, then MC3 aand RyanS did MC2

23:10 UTC SUSHB13 crashed, possible from Marc's work in a close by rack

23:17 UTC ITMX, ITMY WDs tripped, SW, ISI ST1, ISI ST2, and HEPI

23:31 I remeasured MC1 P and confirmed that the Coil Driver swap was unsuccessful, the TF still looked the same. I also ran an undamped OSEM spectra before and after the swap.

Images attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 16:14, Friday 06 March 2026 (89413)
h1susb13 and h1psl0 front end crash from timing errors. New h1ascimc model.

All the h1susb13 models stopped with timing errors this afternoon, followed later by h1psl0's iop model. CER activity was ongoing around this time.

We tried a model restart on h1psl0, which did not work. We rebooted the machine and all started OK.

We tried a reboot of h1susb13, but it got stuck. We then hard-reset the computer via IPMI and all came back OK.

Following the recovery of h1iopsusb13 I untripped the BSC1 and BSC3 SWWDs (the SEI had tripped).

Daniel had a new h1ascimc model to install (no DAQ restart). We stopped h1ascimc, which caused a DAC error on h1iopasc0. So as before, we had to restart all the h1asc0 models to load a new h1ascimc.

H1 SUS
marc.pirello@LIGO.ORG - posted 15:52, Friday 06 March 2026 - last comment - 16:23, Friday 06 March 2026(89411)
MC1 Coil Driver Swap

Per WP13076 we replaced the coil driver for MC1 due to suspicion it was causing MC1 to jump.  The coil driver in question glitched on TOP3.  Referencing D0902810-v12, the chassis in question is U41 on SUS-C4, we exchanged it with a spare.

Old SN - S1001097, New SN - S1100194

Shelia will retest the signal chain to see if this solves the issue.

Comments related to this report
ryan.short@LIGO.ORG - 16:23, Friday 06 March 2026 (89414)

It did not, at least with comparing a before/after transfer function. See alog89409.

H1 SEI
ryan.short@LIGO.ORG - posted 15:19, Friday 06 March 2026 (89408)
Ground Seismometer Mass Position Check - Monthly

FAMIS 38889, last checked in alog89213

There are 12 T240 proof masses out of range ( > 0.3 [V] )!
ETMX T240 2 DOF X/U = -1.784 [V]
ETMX T240 2 DOF Y/V = -1.829 [V]
ETMX T240 2 DOF Z/W = -1.089 [V]
ITMX T240 1 DOF X/U = -2.403 [V]
ITMX T240 1 DOF Z/W = 0.447 [V]
ITMX T240 3 DOF X/U = -2.587 [V]
ITMY T240 3 DOF X/U = -1.122 [V]
ITMY T240 3 DOF Z/W = -3.098 [V]
BS T240 3 DOF Z/W = -0.34 [V]
HAM8 1 DOF X/U = -0.301 [V]
HAM8 1 DOF Y/V = -0.439 [V]
HAM8 1 DOF Z/W = -0.746 [V]

All other proof masses are within range ( < 0.3 [V] ):
ETMX T240 1 DOF X/U = -0.045 [V]
ETMX T240 1 DOF Y/V = -0.067 [V]
ETMX T240 1 DOF Z/W = -0.059 [V]
ETMX T240 3 DOF X/U = -0.048 [V]
ETMX T240 3 DOF Y/V = -0.118 [V]
ETMX T240 3 DOF Z/W = -0.075 [V]
ETMY T240 1 DOF X/U = 0.032 [V]
ETMY T240 1 DOF Y/V = 0.17 [V]
ETMY T240 1 DOF Z/W = 0.232 [V]
ETMY T240 2 DOF X/U = -0.109 [V]
ETMY T240 2 DOF Y/V = 0.203 [V]
ETMY T240 2 DOF Z/W = 0.034 [V]
ETMY T240 3 DOF X/U = 0.244 [V]
ETMY T240 3 DOF Y/V = 0.057 [V]
ETMY T240 3 DOF Z/W = 0.108 [V]
ITMX T240 1 DOF Y/V = 0.293 [V]
ITMX T240 2 DOF X/U = 0.134 [V]
ITMX T240 2 DOF Y/V = 0.244 [V]
ITMX T240 2 DOF Z/W = 0.238 [V]
ITMX T240 3 DOF Y/V = 0.115 [V]
ITMX T240 3 DOF Z/W = 0.065 [V]
ITMY T240 1 DOF X/U = 0.02 [V]
ITMY T240 1 DOF Y/V = 0.142 [V]
ITMY T240 1 DOF Z/W = 0.012 [V]
ITMY T240 2 DOF X/U = 0.018 [V]
ITMY T240 2 DOF Y/V = 0.239 [V]
ITMY T240 2 DOF Z/W = 0.112 [V]
ITMY T240 3 DOF Y/V = 0.043 [V]
BS T240 1 DOF X/U = -0.06 [V]
BS T240 1 DOF Y/V = -0.286 [V]
BS T240 1 DOF Z/W = 0.208 [V]
BS T240 2 DOF X/U = 0.069 [V]
BS T240 2 DOF Y/V = 0.117 [V]
BS T240 2 DOF Z/W = 0.022 [V]
BS T240 3 DOF X/U = -0.132 [V]
BS T240 3 DOF Y/V = -0.259 [V]

There are 2 STS proof masses out of range ( > 2.0 [V] )!
STS EY DOF X/U = -4.566 [V]
STS EY DOF Z/W = 2.375 [V]

All other proof masses are within range ( < 2.0 [V] ):
STS A DOF X/U = -0.576 [V]
STS A DOF Y/V = -0.668 [V]
STS A DOF Z/W = -0.645 [V]
STS B DOF X/U = 0.124 [V]
STS B DOF Y/V = 0.96 [V]
STS B DOF Z/W = -0.355 [V]
STS C DOF X/U = -0.947 [V]
STS C DOF Y/V = 0.891 [V]
STS C DOF Z/W = 0.653 [V]
STS EX DOF X/U = 0.578 [V]
STS EX DOF Y/V = -0.515 [V]
STS EX DOF Z/W = -0.367 [V]
STS EY DOF Y/V = 1.166 [V]
STS FC DOF X/U = 0.155 [V]
STS FC DOF Y/V = -1.181 [V]
STS FC DOF Z/W = 0.633 [V]
 

H1 IOO (IOO, ISC)
masayuki.nakano@LIGO.ORG - posted 15:17, Friday 06 March 2026 (89407)
JAC length motion estimated

Estimated Length Motion of the JAC

The calibrated error signal and feedback signal spectra were measured to estimate the free-running length motion of the JAC. Below the unity gain frequency (UGF = 400 Hz), the feedback signal represents the cavity length motion, while above the UGF the error signal represents the motion. The estimated length noise is well below the design assumption and does not appear to limit the JAC performance.

Method

The spectra of the calibrated error signal and feedback signal were measured to estimate the free-running length motion of the JAC.
- Below the UGF (400 Hz), the feedback signal represents the cavity length motion.
- Above the UGF, the error signal represents the cavity length motion.
Around 400 Hz, the spectrum appears slightly inflated because of the phase bubble, but the actual noise level is expected to be approximately flat in that region.

Observed Features in the Spectrum

- Above 30 Hz
Above 30 Hz, the spectrum becomes flat. This is likely not real cavity motion, but instead electronics noise from the readout chain. A more careful calculation is needed to identify the exact source, but it is most likely the photodiode or the ADC.
Since the incident power is currently 1 W, this noise floor is expected to decrease if the input power is increased.

- Below 10 Hz
Toward DC, the spectrum rises approximately with an f-3 slope. This is interpreted as drift in the PZT control signal caused by temperature drift.
Therefore, the actual cavity length variation at low frequencies is expected to be smaller than what appears directly in the measured spectrum.

- Estimated Length Noise
Looking at the spectrum around 10 Hz, where the above effects are not expected to dominate, the cavity length noise is estimated to be approximately 2 × 10-14 m/rtHz.
In the design, the cavity length motion was conservatively assumed to be 1 × 10-12 m/rtHz at 10 Hz.
Therefore, the measured result is well below the design assumption, indicating that the loop design and the JAC do not introduce problematic intensity noise or phase noise.

Sanity Check with the FSS Unlocked

As a sanity check, the same measurement was repeated with the FSS unlocked. In this measurement, the vertical axis was converted into laser frequency noise by multiplying the calibrated length signal by
FSR / (lambda / 2)
where the free spectral range is given by
FSR = c / L
with c the speed of light and L the cavity round-trip length. For the JAC, L = 2.02 m.

The resulting spectrum, shown in the second plot, is approximately 100 Hz/rtHz at 100 Hz. This is consistent with the typical frequency noise of the NPRO laser.
This also confirms that the JAC is sufficiently quiet compared with the NPRO noise level.

Images attached to this report
H1 IOO
sheila.dwyer@LIGO.ORG - posted 14:31, Friday 06 March 2026 (89405)
IMC refl phasing looks good

Today we have been having some difficulty in locking the IMC, which might be related to MC1 an MC1 suspension problem.  

With the mode cleaner aligned and flashing, we checked the IMC REFL RF24 I and Q signals to check that the PDH phase looks OK, it does.  This is a similar amplitude to what we saw in trends from a time when the IMC was locking, here.  

Images attached to this report
H1 CDS
david.barker@LIGO.ORG - posted 13:53, Friday 06 March 2026 (89404)
h1sush6 front end build

WP13043 h1sush6 front end install

Daniel, Erik, Fil, Jonathan, Dave:

The new h1sush6 front end system is running with its full card complement and a basic IOP model.

On Wednesday afternoon we got the computer booting and seeing most of its chassis, detailed in alog 89376

The outstanding issues were: timing card was not receiving a timing signal, 4th Adnaco BP was not seen.

On Thursday we tracked the fibre issues to a not-quite-seated MTP on the MSR's MER patch (port 3). Once this fibre was reseated correctly the timing card received its signal and the 4th Adnaco BP was seen.

At this point I built up the IO Chassis with the correct card layout, using the ADC and 16bit-DACs provided by the BHD group. We supplied the Interface cards and ribbon cables from stock.

As of end-of-business Thursday the IO Chassis was almost complete, I had miscounted the 16bit-DACs and we were one card short. I built h1iopsush6 with this partial layout and we got the model running.

Friday lunch time I installed the 5th 16bit-DAC and added it to h1iopsush6. The system is now complete as-per drawings G2301306

H1SUSH6 IO Chassis Layout

A1-1 LIGO Timing Card S2101110   A3-1 16AO16-DAC4 250611-24 S2500772 --- S1500324
A1-2 empty   A3-2 empty
A1-3 16AI64-ADC0 S2500754 --- S1102353   A3-3 empty
A1-4 16AO16-DAC0 110419-25 --- S1900099   A3-4 empty
         
A2-1 16AI62-ADC1 210128-39 S2500747 --- S1102355   A4-1 6464Contec-BIO0
A2-2 16AO16-DAC1 250611-03 S2500773 --- S1102283   A4-2 empty
A2-3 16AO16-DAC2 250911-07 S2500768 --- S1900098   A4-3 empty
A2-4 16AO16-DAC3 25-611-10 S2500770 --- S15000314   A4-4 empty

 

H1 ISC
sheila.dwyer@LIGO.ORG - posted 13:38, Friday 06 March 2026 (89401)
ISCT1 ALS SHG alignment

This morning I went to the bifurcated laser hazard and adjusted alignment of the ALS SHG path on ISCT1, picking up where Jenne left off yesterday.  Drawing here

The IR beam was reasonably centered on the bottom periscope mirror and the 1" steering mirror right after it, it was clipping on the 1" BS used for the monitor PD.  I yawed the 1" mirror right after the periscope to reduce the clipping, seen on a card right after the BS.  Then I continued to yaw that first steering mirror to get a beam transmitted through the SHG, as you translate the beam you can see lots of glints transmitted which aren't the actual beam.  As soon as there was full beam going through the SHG, the green beam was alinged onto the COMM A broadband PD, so I didn't move any steering optics other than the first one.  

The monitor diodes do not agree with the power levels that they showed before the table move.  The IR one shows 6 times more power than before the table move, which makes me think the diode may not have been well aligned at the end of O4.  The beam in reflection off that first beamsplitter has two lobes visible on a card, possibly the front and back surfaces of the BS.  The beam was clearly missing the green power monitor PD, (before photo), I moved the diode to center, but the beam is still too low.  Nearly a year ago TJ and I wrote that this diode wasn't working, 84558, it does but seems to have been working since June 9th 2025, a day when the table was realigned as described in 84900. Since this alignment puts the beam onto the COMM diode, which is quite far from the SHG, I think it must be similar to previous alignments, and we probably need to adjust the height of this diode.

The power on the COMM A PD is about 300 counts, similar to what it was in O4.  So, this is probably a good enough alignment for us to proceed with locking with.  

Images attached to this report
H1 IOO (IOO, ISC)
masayuki.nakano@LIGO.ORG - posted 13:28, Friday 06 March 2026 (89402)
JAC LSC loop design

Summary

The JAC length servo was designed to set the unity gain frequency (UGF) at 400 Hz. Additional low-frequency boost was implemented to improve suppression below 50 Hz. The open loop gain (OLG) was measured and compared with the servo model.

Details

 

Images attached to this report
H1 IOO (IOO, ISC)
masayuki.nakano@LIGO.ORG - posted 12:44, Friday 06 March 2026 (89399)
JAC error and feedback signal calibration

Summary

The calibration of the PDH error signal and the feedback path was derived using the Guardian-based signal normalization. The normalized PDH error signal allows the optical gain to be calculated analytically. The transfer function from L_SERVO_OUT to the cavity length actuation was measured and modeled, separating the optical gain and the PZT actuator response.

Details

Normalized PDH error signal

With the Guardian normalization, the PDH error signal at L_SERVO_IN1 can be written as
V = x / (1 + x^2)
where V is the signal at L_SERVO_IN1, and
x = l / HWHM
where l is the cavity length fluctuations and HWHM is the cavity half-width at half-maximum.

Using the finesse F, the cavity HWHM is
HWHM = lambda / (4F)

At the lock point (x = 0), the slope of the error signal is
dV/dx = 1

Optical gain

Therefore, the optical gain is
dV/dl = dV / d(x * HWHM) = 4F / lambda

Using F = 125, lambda = 1064e-9 m, the optical gain becomes
dV/dl = 4.70e8 cnts/m

Error signal calibration

To convert the signal at L_SERVO_IN1 to cavity length, we apply the inverse of the optical gain.
Calibration factor = 2.128e-9 m/cnts

Plant measurement

After locking the cavity with a provisional filter, the transfer function from L_SERVO_OUT to L_SERVO_IN1was measured and treated as the plant (see attached plot).
Since this plant includes both the optical gain dV/dl and the PZT actuator response, L_SERVO_IN1 was converted into meters using the calibration factor above before the measurement.
Also, the servo output is calibrated in V (and converted into cnts at the PZT_DRV filter). That means, the measured plant represents the transfer function from the PZT driver input to the actual cavity length actuation with the Unit of m/V

Plant model and actuator calibration

The optical gain and PZT actuator response are implemented in the servo model as FM9 and FM10 of L_SERVO.
In addition, a zpk(-800, 800) filter is included to emulate the phase delay.
The comparison between the model (FM9*FM10) and the measured plant (uncalibrated) is shown in the second plot. This response includes the PZT driver transfer function. That has two poles at 1 Hz and 400Hz, and one zero at 10 Hz. The DC gain estimated from the measured TF is 2.57 nm/V. This is comparable to the value measured with the scan of the JAC (2.97 nm/V).

Error signal normalization consistency

The error signal at JAC-L_SERVO_IN is normalized by the power at output of JAC_REFL_A_RF43. Therefore, once the guardian normalization procedure has been executed, the same calibration factor should remain valid.

 

Images attached to this report
H1 SUS
sheila.dwyer@LIGO.ORG - posted 11:55, Friday 06 March 2026 - last comment - 15:33, Friday 06 March 2026(89398)
MC1 suspension large shift not from requested drive

I was adjusting the alignment of MC1 pitch earlier, when the suspension had a large alignment shift that wasn't caused by a change in the requested drive to the DACs.  The attached screenshot shows my requested alignment shifts in opticalign, and the resulting change in the requested DAC counts on T2 + T3.  There is a sudden 420 urad jump in the osem readbacks from pitch, and a 133 urad jump in roll at the first time cursor.  The master outs only see a few cycles on oscillations, which probably is from the damping loops responding to the jump, but they settle to the same requested drive as before the jump.  The jump seems to be real as the IMC flashes disappeared at this time.  

Images attached to this report
Comments related to this report
ryan.crouch@LIGO.ORG - 14:28, Friday 06 March 2026 (89400)

I ran an undamped no alignment offsets suite of transfer functions for MC1, and MC3, the state of HAM2 at the time was HEPI locked, ISI isolated. MC1 Pitch was the worst looking TF with multiple extra peaks. Looking at the OSEM signals for MC1 T3 shows a questionable drop during the time Sheila was looking at.

MC1:

/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/MC1/SAGM1/Data/

2026-03-06_1930_H1SUSMC1_M1_WhiteNoise_P_0p02to50Hz.xml
2026-03-06_1930_H1SUSMC1_M1_WhiteNoise_R_0p02to50Hz.xml
2026-03-06_1930_H1SUSMC1_M1_WhiteNoise_V_0p02to50Hz.xml
2026-03-06_1930_H1SUSMC1_M1_WhiteNoise_Y_0p02to50Hz.xml

2026-03-06_1930_H1SUSMC1_M1_WhiteNoise_T_0p02to50Hz.xml
2026-03-06_1930_H1SUSMC1_M1_WhiteNoise_L_0p02to50Hz.xml

MC3:

2026-03-06_2130_H1SUSMC3_M1_WhiteNoise_L_0p02to50Hz.xml
2026-03-06_2130_H1SUSMC3_M1_WhiteNoise_P_0p02to50Hz.xml
2026-03-06_2130_H1SUSMC3_M1_WhiteNoise_R_0p02to50Hz.xml
2026-03-06_2130_H1SUSMC3_M1_WhiteNoise_T_0p02to50Hz.xml
2026-03-06_2130_H1SUSMC3_M1_WhiteNoise_V_0p02to50Hz.xml
2026-03-06_2130_H1SUSMC3_M1_WhiteNoise_Y_0p02to50Hz.xml

Images attached to this comment
ryan.short@LIGO.ORG - 15:07, Friday 06 March 2026 (89406)

I ran the same set of transfer functions for MC2 while Ryan did the other MCs; nothing suspicious here.

Templates' location: /ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/MC2/SAGM1/Data/2026-03-06_2145_H1SUSMC2_M1_WhiteNoise_{P,R,T,V,Y,L}_0p02to50Hz.xml

Images attached to this comment
ryan.crouch@LIGO.ORG - 15:33, Friday 06 March 2026 (89409)

Remeasuring MC1_P after Marc swapped the coil driver reveals that the coil driver was not the issue.

Images attached to this comment
H1 ISC (CDS, INS, ISC)
keita.kawabe@LIGO.ORG - posted 18:13, Thursday 05 March 2026 - last comment - 10:41, Friday 06 March 2026(89389)
JAC WFS DC not going into the right channel. RF channels are working. (Jennie, Jason, Keita)

Jason and Jennie found that there's no digital signal coming out of JAC WFS DC no matter what.

I and Jason went to the floor, I flipped the gain switch of the WFS interface from low to high, but there was absolutely no response from any of H1:JAC-WFS_[AB]_SEG[1234]_INMON.

We confirmed that interface cables are connected from the IOT1 to the WFS interface (ISC-R1 slot U10), the interface is connected to the AA (ISC-C1 slot U30), and the AA chassis is connected to ADC5. These all agree with D1900511 floor wiring diagram. I even disconnected the cable coming to AA (ISC_373) and WFS DC signals didn't respond.

It seems that there's discrepancy between the floor wiring diagram and the model which was made according to T1100472 where JAC WFS DC signals are routed to ADC3 (see attached).

Good news is that RF channels are working.

As of now, the light is hitting WFSA and I was able to phase RF using an excitation injected into JAC PZT while JAC itself was locked. WFSB is dark now. 

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 18:18, Thursday 05 March 2026 (89390)

Jennie W, Jason O

 

Starting about 1.45pm, we went to check the alignment on the table after getting the ok from vacuum to let light into HAM1.

We could see light getting through the periscope but Jason tweaked the upper mirror to get the beam centred on the lower mirror.

Then we centred the mirrors M5, M7, BS3 and M8 to centre on each mirror and then align onto the REFL PD. The shutter and PBS1 were not in the beam.

We had some problems aligning the HWP/PBS combo while HAM1 was still in air but today Jason managed to align the PBS and the angle of the half-wave plate such that the power to the PD is maximised.

The beam dump for the off-polarisation is also aligned. We had to tweak BS3 and M8 to re-align onto the PD.

After this we went onto aligning the beam onto the WFS and placing beam dumps for the beams reflected from each QPD.

Even when WFS A looked well-aligned by checking with the card, we could not see signals on WFS A quadrants. See explanation above in Keita's log.

Jason left WFS B mis-aligned as the beam tilts up slightly and so we could not catch the reflected beam with the beam dump.

We didn't want to fix this as Masayuki is working on tuning the JAC length servo, so we don't want to move M7 now.

jennifer.wright@LIGO.ORG - 18:27, Thursday 05 March 2026 (89391)

According to T1100472 ADC 5 channels 17-24 should be something called  "LO_A_DC" and "LO_B_DC" but I can't find these ADC inputs hooked up to anything in h1ascimc or h1asc.

We can do a model restart to fix this (ie. hook up the JAC WFS DC readouts to channels on ADC5 instead of ADC3) on Monday assuming that is ok with Daniel/Dave.

keita.kawabe@LIGO.ORG - 08:54, Friday 06 March 2026 (89395)

JAC and LO WFS DC assignments are swapped between the channel assignment document (T1100472) and the wiring diagram (D190511).

Channel assignment: JAC WFS DC = ADC3, DB9_3 and DB9_4. LO WFS DC = ADC5 DB9_5 and DB9_6.

Wiring diagram: LO WFS DC = ADC3, DB9_3 and DB9_4. JAC WFS DC = ADC5 DB9_5 and DB9_6.

Assuming that the above is correct, you can do one of two things.

1. Swap cables on the AA chassis and update the wiring diagram.

2. Change the model and update the channel assignment.

daniel.sigg@LIGO.ORG - 10:41, Friday 06 March 2026 (89397)
T1100472 updated
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 00:20, Tuesday 03 March 2026 - last comment - 15:54, Friday 06 March 2026(89326)
HAM1 Pump Down

(Travis S., Gerardo M.)
Pumpdown was started this morning, sorry no blowdown data.  By the time I remembered to take the reading there was no more pressure inside the chamber.  Travis connected all of the equipment while I wrestled with the viewport inspection.
Pumpdown was started without an issue.  We started with the scroll pump from the super sucker 500, the scroll pump of the SS500 cart brought the pressure down to 3.7x10-01 Torr, then we switched the cart to start pumping with the turbo pump/scroll pump.  We had one trip of the system, usually due to a time out, but it only took a restart of the turbo pump and the pumpdown to continue.
Unfortunately the pumpdown had to be stopped at 8:00 PM local time, with a pressure of 4.0x10-05 Torr, at that pressure the set points had not activated on the SS500 cart controller, the set points are hard set and can't be modified from the SS500 controller screen.
Pumpdown will be restarted tomorrow morning.

Note, on the attachment, regarding the screens, no set points for channel 1.

Images attached to this report
Comments related to this report
gerardo.moreno@LIGO.ORG - 15:54, Friday 06 March 2026 (89410)VE

Pumpdown Update.

Last Tuesday, HAM1 pumpdown was stopped for a few minutes to move the SS500 cart, the cart was moved away from the -Y door to open up space for the table. After the relocation of the SS500 cart the pumpdown for HAM1 was restarted.

On Thursday we assisted Filiberto with the installation of a "split" cabling for the high voltage interlock at HAM1 gauge.  Attached is a photo of the screen of the SS500 cart and its response to loosing gauge signal, the signal cable was removed from the vacuum gauge to install the "split".  I set valve control to "open" and the override to "override" on the SS500 cart to prevent it from tripping, it worked.  BTW, when the cart reported the Error-S it cleared the setpoints.  Once the cable was restored and we had a signal at the SS500 cart, the setpoints were restored manually.

Pumpdown continues, pressure is currently at 2.77x10-06 Torr, attached is a trend of the pumpdwon pressure.

Images attached to this comment
LHO VE (VE)
travis.sadecki@LIGO.ORG - posted 15:18, Friday 27 February 2026 - last comment - 16:01, Monday 09 March 2026(89308)
HAM1 +Y door installed

The HAM1 +Y door was installed today.  Installation went smoothly and the O-ring cooperated by staying in place today.  I will note that torquing the door bolts is becoming more of an issue due to wrench access being increasingly limited near the 5-way cross/feedthru areas now that all the in-air cables have been attached.

The HAM1/HAM2 annulus system is being pumped by 2 turbos/aux carts on the -Y side of the chambers.  The mysterious ~order-of-magnitude pressure readout difference at the two carts on the shared annulus system is again present, as was noted the last time this annulus system was pumped down.  Not an issue per se, just some vacuum phenomena that leaves us scratching our heads. 

Comments related to this report
gerardo.moreno@LIGO.ORG - 16:33, Friday 06 March 2026 (89415)VE

Vent Recovery Update

Annulus ion pumps for HAM1 and HAM2 continue to pumpdown, we closed the isolation valves for both of them, but the ion pumps were not able to maintain pressure.  On 3/4/26 valves were opened again to continue using the aux carts, and yesterday (3/5/2026) we isolated the aux cart for HAM2 only, all of that can be seen on the attached plot.

Aux carts will be left pumping and annulus systems are going to be evaluated next Monday.

Images attached to this comment
jordan.vanosky@LIGO.ORG - 16:01, Monday 09 March 2026 (89440)

I isolated the HAM1 aux cart (HAM2 was isolated on Friday) from the annulus this morning. The IP current only rose to ~4.7mA before turning around showing it could hold the annulus pressure.

HAM1/2 annulus aux carts were shut down and decoupled this afternoon. The SS500 cart remains at HAM1 and will continue to pump until the pressure is low enough to valve in the HAM1 ion pump.

Images attached to this comment
LHO VE
jordan.vanosky@LIGO.ORG - posted 13:37, Thursday 19 February 2026 - last comment - 16:07, Friday 06 March 2026(89199)
Leak Checking of the HAM7 +Y door ports and the Relay Tube

Now that HAM7 is at sufficient pressure (6.4E-7 Torr as of this post) we helium leak checked the two re-installed 12" CF blanks removed for table locking and the entire relay tube assembly which had been removed for the viewport adapter on HAM5.

The Helium background was unstable, so we turned on the HAM7 cleanroom (on 2/18 afternoon) to help flush the area of helium overnight.

Today the background stabilized around 1.5E-10 Torr-l/s, so we were able to continue leak checking. No significant helium signal observed above the leak detector background which hovered between 1.5&2E-10 Torr-l/s during leak checking. 

HAM7 continues to pump down with the turbopump, next steps are chamber RGA scans.

Images attached to this report
Comments related to this report
jordan.vanosky@LIGO.ORG - 16:48, Thursday 26 February 2026 (89292)

HAM7 RGA scans and scan information posted to T2600064. RGA is dirty, but the chamber comparison to last pumpdown (Mar 2024) is posted.

gerardo.moreno@LIGO.ORG - 06:32, Friday 27 February 2026 (89296)VE

Late entry.

Last Tuesday, 2/24/2026, HAM7 ion pump was introduced to the chamber's main volume, and currently the chamber is pumped with both; the ion pump and the turbo pump.  Attached is a trend snap of the chamber's pressure since the introduction of the ion pump.

Images attached to this comment
gerardo.moreno@LIGO.ORG - 16:07, Friday 06 March 2026 (89412)VE

Vent Recovery Update

On Wednesday (3/4/26) the turbo pump was isolated from HAM7 to leave the ion pump take care of the pumpdown. 

Since the ion pump was able to take of the pumpdown on its own, on Thursday the SS500 cart then was turned off, and eventually detached from the turbo pump at HAM7.  HAM7 is back to normal operations.

Images attached to this comment
Displaying reports 1781-1800 of 88643.Go to page Start 86 87 88 89 90 91 92 93 94 End