TO DO - (PARALLEL) Log in the information about the pressure changes and the plan for the estimator. - Gear up to take a measurement for OSEM symmetrization on SR3. + Remember to grab the vertical/etc. positions of the OSEMS + Remember to Isolate the ISI + Remember to change the OSEM damping gains + Remember to grab the drive coil compensation and change it to 1 + Remember to monitor the current pressure state - (PARALLEL) Log in the information about the PR3 changes in calibration after the vent. + Grab the vertical positions of the OSEMs for both measurements + Compare the numbers that we are getting. ------------------------------------------- - OSEM POSITIONS (at M1 DAMP) + L 13 + T -6 + V -35 + R -582 + P -309 + Y -515 - ISOLATIONG HAM5 ISI + HEPI ISO GAIN slider set to zero + ISOLATED state requested for HAM5 - Turned Coil Driver compensation to state 1 - OSEM POSITIONS (at M1 DAMP) (rev) + L 14 + T -7 + V -35 + R -582 + P -313 + Y -515 - GPS Time 1429034767 - Pressure on PT120A: 1.35 Torr - Starting DTT - Changed gains in the OSEMS to -0.1 - ISI to M1 Measurement start is approx 19:00 UTC on April 18th + ABORTED - The filters on SUSPOINT DRIVE on HAM5 were all off -Turning filters ON - ISI to M1 Measurement start is approx 19:05 UTC on April 18th + Yaw measurement only for now - Changed gains in the Yaw M1 DAMP to -0.1, keeping the rest at -0.5 (nominal for SR3) + Measured ISI L, T , V, and Y + Measured R and P - Grabbing M1 TFs at 22:02 UTC + Got all transfer functions Committed the transfer functions to revision 12256 All of the changes were reverted by Brian and me at around 00:30 UTC.